VRA Safety and Regulatory Information

Applies to the 3D Toy VRA (XM-5000 Series 5) and HD Video VRA (VDS-3000 Series 3), RF and IR. Revision A, November 24, 2026.

This page holds the full safety and regulatory information for Pehratek VRA systems. A short version is printed on the Easy Install Guide packed with your system. For a free paper copy of this page, call 855-327-8440 or email info@pehratek.com.

Intended use

Pehratek VRA systems provide the visual or audiovisual reinforcement used during Visual Reinforcement Audiometry (VRA), a pediatric hearing-assessment technique. During a test, separate audiometric equipment presents a sound, the child turns toward it, and the clinician uses the remote to turn on a toy, lights, or video on that side. An optional center unit helps bring the child's attention back to the front before the next sound.

Intended users. The systems are for use by audiologists and other hearing-care professionals trained in VRA, and by staff working under their direction. They are not for use by patients, parents, or children.

Intended environment. Normal indoor clinical settings: audiology, pediatric and ENT clinics, hospital audiology departments, school and university clinics, research and training settings, and sound booths with their control areas.

What the system does not do. The Pehratek VRA system is not the audiometer. It does not generate the calibrated test signal, measure hearing thresholds, or calculate or display a test result. It does not connect to the patient, needs no acoustic calibration, and stores no patient records or test results. Audiometers and sound-field loudspeakers stay under their own manufacturers' instructions and calibration requirements.

Clinical technique, patient conditioning, diagnosis, and interpretation of results are the responsibility of the clinician.

Safety information

The words below mean the following. WARNING: a hazard that could cause injury. CAUTION: a condition that could damage the product or make it work incorrectly. NOTICE: other important information.

Read the Easy Install Guide before you mount or connect anything, and keep it with the system. Take damaged equipment out of use until it has been checked.

Where not to use the system

  • Wet locations, or locations with uncontrolled condensation
  • Outdoors, or anywhere exposed to weather
  • Explosive or flammable environments
  • Oxygen-enriched environments, unless separately evaluated and approved
  • MRI scanner rooms or controlled magnetic-field zones
  • Locations with heavy dust, vibration, heat, cold, or corrosive chemicals

Mounting

  • WARNING: An insecurely mounted unit may fall and cause injury or equipment damage. Mount every unit with the hardware supplied and with fasteners suited to your wall type.
  • Confirm that the wall, booth panel, speaker cabinet, or arm can hold the unit. If you are unsure, ask a qualified installer.
  • For HD Video monitors, follow the instruction sheet packed with the arm bracket, including its limits on wall type and weight.
  • Follow your facility's rules for drilling into sound booth walls.
  • Inspect the mounting hardware after installation and after any move, impact, or adjustment. Tighten loose hardware, and replace any that is bent, cracked, stripped, or missing, before the unit goes back into use.
  • Route cables so they do not pull on a mounted unit. Do not use a power or receiver cable as a safety tether.

Cables

  • WARNING: Do not place cords where a child can pull, wrap, or become entangled in them. Cables within a child's reach are a strangulation and tripping hazard.
  • Route cables away from walkways and from areas patients can reach. Secure excess cable without sharp bends.
  • Do not use damaged cables, and do not splice Pehratek control cables.

Flashing lights and video

  • WARNING: These systems use flashing and moving LED lights and moving video. A small number of people can have a seizure when exposed to flashing lights or certain visual patterns, including people with no history of seizures.
  • Before a session, ask the parent or caregiver whether the child has a history of seizures or photosensitive epilepsy. The clinician decides whether and how to use the system for that child.
  • The remote lets you turn the LED lights off and choose the light pattern, the toy action, and the video.
  • Stop the reinforcement at once if a child, or anyone else in the room, shows signs such as eye or muscle twitching, loss of awareness, disorientation, or convulsions, and follow your clinic's emergency procedure.

Magnets and small parts

Toys in the 3D Toy VRA attach with magnets. Each toy unit has a magnetic base, and the toy customization kit contains loose ring magnets.

  • WARNING: Swallowing hazard. Swallowed magnets can attract each other inside the body and cause serious injury or death. Keep loose magnets, the customization kit, and any toy that has come off its unit out of the reach of children. If you think a child has swallowed a magnet, get emergency medical care at once.
  • WARNING: Implanted and body-worn medical devices. Magnets can affect how some medical devices work. These include pacemakers, implantable defibrillators, programmable shunts, and cochlear implants and their sound processors. Keep loose magnets and magnet-fitted toys at least 6 inches (15 cm) away from any such device. Do not hand a magnet-fitted toy to a child or adult who has one. When in doubt, follow the instructions from the maker of the medical device.
  • Before each session, check that the toy is seated firmly on its base and that no magnet is loose.
  • Store spare magnets in the kit box, away from patient areas.
  • Follow the card in the kit when you fit a magnet to a new toy. Take a toy out of use if its magnet comes loose.

Power

  • CAUTION: The units run from external low-voltage DC power supplies. Use only a power supply supplied or approved by Pehratek for your model. Do not substitute one because its connector or voltage looks the same.
  • Do not use a power supply with a cracked case, bent blades, cut insulation, exposed wire, discoloration, an unusual smell, or signs of overheating.
  • Unplug a unit before cleaning it or moving it.

Remote control batteries

  • The remote uses standard AA batteries. Replace them as a set and follow the polarity marks.
  • Remove the batteries if the remote will sit unused for a long period. Dispose of used batteries under your local rules.

Cleaning

  • Disconnect power first. Wipe with a soft, lint-free cloth lightly dampened with a mild, alcohol-free cleaner suitable for electronics and plastic. For the video screens, use a cleaner labeled for LCD or LED displays, applied to the cloth.
  • CAUTION: Do not spray cleaner onto the equipment, immerse any part, or let liquid enter ports, seams, buttons, or vents. Let everything dry fully before you reconnect power.
  • Do not use alcohol-based cleaners, solvents, acetone, ammonia-based glass cleaner, abrasive cleaners, or scouring pads.

Service

  • WARNING: Do not open the HD Video monitor. It has no user-serviceable parts inside.
  • Users may clean the equipment, replace remote batteries, and inspect mounts and cables. For anything else, contact Pehratek.
  • A service manual for biomedical and clinical engineering staff is available on request.

Radio and electromagnetic compliance

United States (FCC)

RF remote control: FCC ID 2A6D4XM5000. The same RF remote is used with the XM-5000 Series 5 and the VDS-3000 Series 3. The remote is the only part of the system that transmits a radio signal. IR systems use infrared light and do not transmit radio signals.

This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.

Changes or modifications not expressly approved by Pehratek Products, LLC, the party responsible for compliance, could void the user's authority to operate the equipment.

Responsible party: Pehratek Products, LLC, 1620 N Interstate 35E, Ste 305, Carrollton, TX 75006.

Standards

Pehratek products have been tested to the following standards:

  • FCC Part 15.231 (United States, radio)
  • RSS-210 (Canada, radio)
  • EN 300 220 (Europe, radio)
  • EN 301 489-3 (Europe, electromagnetic compatibility)
  • EU Radio Equipment Directive 2014/53/EU, Article 3.2

Pehratek products comply with the EU RoHS and REACH substance requirements.

Test reports, declarations, and certification documents are available from Pehratek on request. Tell us the product model, series, serial number, RF or IR configuration, and the country of installation.

If you suspect interference

  1. Stop testing.
  2. Confirm you are using the right remote and receiver.
  3. Move other electronic equipment farther away where you can.
  4. RF systems: move the RF receiver away from metal objects, and confirm that nearby Pehratek systems are set to different channels. The guide "Changing RF Channels" is on the support page.
  5. IR systems: keep each sensor off the monitor and in a clear line of sight to the remote.
  6. Contact Pehratek if the problem continues.

Disposal

Dispose of electronic equipment and batteries under your local rules. Do not put them in household trash where separate collection is required.

Contact

Pehratek Products, LLC
1620 N Interstate 35E, Ste 305, Carrollton, TX 75006
855-EAR-8440 (855-327-8440) · info@pehratek.com

To report a problem with a Pehratek system, contact us at the number or email above.